A Nonlinear Calibration Method Based on Sinusoidal Excitation and DFT Transformation for High-Precision Power Analyzers
Table 2
100 Hz current measurement comparison.
| AC current (A) signal frequency 100 Hz | Standard value | The old method | The proposed method | Phase A | Phase B | Phase C | Phase A | Phase B | Phase C |
| 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.50 | 0.49 | 0.50 | 0.49 | 0.49 | 0.49 | 0.49 | 1.00 | 0.99 | 0.99 | 0.99 | 0.98 | 0.99 | 0.98 | 5.00 | 5.00 | 4.99 | 5.00 | 4.97 | 4.98 | 4.99 | 10.00 | 10.04 | 9.98 | 10.01 | 9.99 | 9.98 | 10.01 | 25.00 | 25.09 | 24.81 | 25.11 | 24.90 | 24.98 | 24.94 | 40.00 | 38.98 | 39.01 | 38.64 | 39.85 | 40.04 | 39.96 | 55.00 | 52.18 | 51.95 | 52.02 | 54.24 | 54.43 | 54.32 |
|
|