Research Article

A Nonlinear Calibration Method Based on Sinusoidal Excitation and DFT Transformation for High-Precision Power Analyzers

Table 3

500 Hz current measurement comparison.

AC current (A) signal frequency 500 Hz
Standard valueThe old methodThe proposed method
Phase APhase BPhase CPhase APhase BPhase C

0.100.100.100.100.100.100.10
0.500.490.500.490.490.490.49
1.000.980.970.980.980.990.98
5.005.014.995.004.995.014.99
10.0010.049.9710.029.979.9810.01
25.0025.2524.7625.2124.9524.9524.90
40.0038.6838.8938.3539.8840.0239.97
55.0051.8451.7251.8854.3454.4554.40