Research on the Three-Dimensional Power Frequency Electric Field Measurement System
Table 7
Comparison between EHP-50F and 3D power frequency electric field measurement system as function of the distance.
| Distance (m) | Simulation (V/m) | EHP_50F (V/m) | Three-dimensional electric field measurement system (V/m) | Deviation (%) |
| 0.1 | 24708.55 | 24709 | 24689 | -0.08% | 0.3 | 8235.66 | 8236 | 8271 | 0.43% | 0.5 | 4941.28 | 4942 | 4954 | 0.26% | 0.7 | 3529.37 | 3530 | 3513 | -0.46% | 0.9 | 2745.66 | 2745 | 2750 | 0.16% | 1.1 | 2246.5 | 2246 | 2239 | -0.33% | 1.3 | 1899.56 | 1901 | 1896 | -0.19% | 1.5 | 1647.94 | 1647 | 1639 | -0.54% | 1.7 | 1453.57 | 1454 | 1453 | -0.04% | 1.9 | 1300.62 | 1301 | 1295 | -0.43% | 2.1 | 1176.39 | 1177 | 1180 | 0.31% | 2.3 | 1073.86 | 1074 | 1069 | -0.45% | 2.5 | 988.55 | 989 | 988 | -0.06% |
|
|