Abstract

Low-cost PM2.5 sensors are the key to creating high-resolution monitoring networks for localized reporting and increased public awareness. A low-cost PM2.5 sensor using light-scattering technology Public Health Particulate Matter (PHPM) was developed and validated against measurements from a commercial device that reports real-time aerosol concentration and also collects gravimetric samples (DustTrak DRX Aerosol Monitor 8533). Linear regression of measurements from a controlled indoor site and three ambient air sites found that the novel PM2.5 sensor correlated well with the measured PM2.5 concentrations () at the three ambient air locations with PM2.5 concentrations ranging from 48 to 602 μg/m3 measured at 30-minute intervals. The correlation was lower () at the chamber, which used incense as a particulate matter source. The PHPM data of PM2.5 concentrations was using multiply with 2.33. The hope is that the PHPM device can provide an additional tool for participatory in community and increased public awareness about PM2.5 in Thailand. This PHPM device is more suitable for ambient air than the chamber. The PHPM device is inexpensive, portable, and can be charged with a power bank.

1. Introduction

Particulate matter air pollution with an aerodynamic diameter less than 2.5 microns (PM2.5) is recognized as a global health threat that requires prioritized research [1]. While efforts to monitor ambient air quality have increased substantially across the globe, significant gaps remain, with many parts of the world having no regular PM2.5 monitoring [2]. Localized measurements of PM2.5 are important due to the high variability in ground level air pollution caused by complex interactions of topography [3], air patterns [4], and emission sources [5]. Two major barriers, however, to widespread implementation of PM2.5 monitoring are the cost and technical requirements of high-accuracy equipment that relies on filtration and gravimetric mass determination [6]. Furthermore, gravimetric methods typically generate data with 24-hour measurement periods, while higher temporal resolution is preferable for real-time monitoring, highly sensitive for the detection of glucose concentrations in human bodies [7], and biosensor has been proposed for detection of uric acid (UA) in human serum with a highly sensitive and selective optical fiber-based enzymatic [8]. In response, light scattering techniques (e.g., Chowdhury et al. [9]) have been developed that build the concept of optical particle counters and Mie’s theory [10]. These light-scattering sensors are inexpensive and easy to operate.

Several low-cost particulate matter sensors have been developed and validated across the world [6, 9]. These sensors have become critical tools for providing localized air quality readings and raising awareness of particulate matter pollution [11]. Advances in computing power and power consumption have made low-cost sensing devices an important supplement to the granular network of regulatory stations established around the globe and help to engage local communities with air quality data [12]. Generally, these low-cost devices can quite accurately measure PM2.5 concentrations under most conditions [13, 14]. However, the accuracy of such low-cost air sensors has been questioned, and the need for calibrating these devices is critical [15]. However, given the knowledge gaps, the aim of this study was to focus on estimating PM2.5 concentrations using low-cost PHPM sensors in real-world situations. Therefore, calibrating and validating low-cost PM2.5 sensors built on light-scattering technology is necessary before implementation in the community. This similarity metric can also guide the adoption of low-cost monitors and/or data selection in future field research as well as spatiotemporal [16]. While PM2.5, which results from vehicle exhaust, biomass burning, and industrial emissions, has received substantial publicity and attention in Thailand [17], high-resolution monitoring of PM2.5 is still limited in Thailand [18]. Widespread local data, however, is important. Attitudes, awareness, and risk perception affect the public’s proenvironmental behaviors that influence PM2.5 [19, 20].

Thus, increasing awareness and publicity of PM2.5 concentrations at the local level is important. Accordingly, the objectives of this research were to develop a simple, low-cost device for real-time reporting of PM2.5 concentrations in the ambient air and to calibrate the device by determining the relationship between the device’s readings and a standardized commercial aerosol sensor. This device is intended for widespread implementation at local health centers to raise awareness about PM2.5 at the local level, as access to real-world PM2.5 data is essential for more effective and safe assessment of people’s health risks and self-protection.

2. Materials and Methods

2.1. Instruments

A low-cost PM2.5 sensing device Public Health Particulate Matter (PHPM) was developed for use in monitoring ambient air PM2.5 concentrations at local health stations or other local agencies (Figure 1(a)). The PHPM uses the light scattering technique to measure particles suspended in the air. A red laser illuminates through a cavity with active ventilation provided by a fan. A photodiode measures the scattered light to determine the mass of particles according to Mie’s theory. These readings were then converted to PM2.5 concentrations by researcher. The estimates from the PHPM device were compared against a standardized commercial product, the DustTrak DRX Aerosol Monitor 8533 (TSI International, Shore-view, Minnesota, USA) (Figure 1(b)). This device can provide real-time aerosol concentration data, while simultaneously measuring both mass and size fraction. While also using a light-scattering laser photometer, it also collects gravimetric samples. These DustTrak monitors have been used to validate or calibrate several low-cost devices for both indoor and outdoor aerosol measurement [21, 22]. The estimates from the PHPM device were compared with the commercial sensors with light scattering technology which include SI, SN, SV, SC, and CKDPM. All commercial devices are shown in Tables 1 and 2.

2.2. Sampling Locations

The two instruments were set up together at four locations from December 2020 to February. Considering human activities in each sampling area to represent each site for reporting measurement results to cover the most residential areas. The first location (chamber) was inside a control room, in which incense was burned to provide a source of aerosols. The second location (University campus) was located on the campus of Khon Kaen University (16.4706°, 102.8255°) (Figure 2). The third location (urban center) was in downtown Khon Kaen, located near high-traffic roads and markets (16.4287°, 102.8354°). The fourth location (rural industrial) was located in a primarily agricultural area but in the vicinity of a sugar cane factors. Data from each instrument was logged every 30 minutes.

2.3. Sampling Method

The concentration of PM2.5 samples were collected for 12 days, with data being recorded every 30 minutes from 8:00 a.m. to 8:00 a.m. the next day for 24 hours. Collect dust samples with a size of not more than 2.5 microns (PM2.5) by installing a collector. Dust samples for both the High-Volume Air Sampler, 5 types of dust measuring instruments (commercially sold brands) and simple dust measuring instruments up to 2.5 μm that were developed from December 2020 to January 2021 for 24 hours from 8:00 a.m. to 8:00 a.m. of the next day for 24 days. Methods for collecting dust samples with a size not exceeding 2.5 μm (PM2.5) by a High-Volume Air Sampler. Preparing filter paper, glass fiber filter is baked to remove moisture for 24 hours. After baking, put the filter paper in the dehumidifier cabinet for at least 24 hours. Then, put the filter paper in the filter head 1 sheet per time used sampling. When the sample collection is complete, the filter paper is placed in the dehumidifier for at least 24 hours and then weighed with a fine balance with 5 decimal places before and after the collected samples were to determine the weight of the dust particles on the filter paper. Method of collecting dust samples particulate matter up to 2.5 μm (PM2.5) by DustTrak 8533, five types of dust measuring instruments (commercially sold brands), and a simple developed dust measuring device up to 2.5 μm. The method of collecting samples of these instruments is simple.

The machine will suck air into the system and analyze the data and display results immediately on the display screen, which will take samples every 30 minutes from 8:00 a.m. to 8:00 a.m. of the following day from December 2020 to January 2021 and averaged by using standard 24 h. Analyzing fine particulate matter PM2.5 by gravimetric method by weighing the dust filter paper with a size of not more than 2.5 μm (PM2.5) before and after, samples were taken, and the weight difference was determined to calculate the concentration of fine particulate matter using the formula as follows: where filter paper weight after sampling (g)

presampling paper weight (g)

volume of air at standard conditions (cubic meter unit)

convert grams to micrograms

Source: Office of Air Quality and Noise Management, Pollution Control Department, Thailand.

2.4. Data Analysis

Pearson’s correlation coefficient, , was computed to measure the association between the PM2.5 estimates from the PHPM and measurements from the DustTrak 8533 for each sampling location independently. Linear regression was used to calibrate the readings of the PHPM to the DustTrak 8533 readings. All data analysis was completed in STATA version 15 (copyright of Khon Kaen University). Simple linear regression was used to describe the relationship between the two variables and to make predictions. By constructing an equation showing a linear relationship between two variables from the linear equation as follows: or , where value is the concentration of PM2.5 obtained from the PM2.5 measuring device. A simple developed and is the reference DustTrak 8530, where or is the -intercept constant and or is the slope or coefficient, where is 0 to 1, and approaching 1, the regression equation can explain a lot of the change. The data were analyzed from the sampling results of dust concentrations up to 2.5 microns by statistical methods, using Pearson correlation coefficient between the two variables. The estimation equation for Pearson correlation coefficient is given as follows: with where is the sample covariance between two random variables and that are normally distributed with means and standard deviations and , respectively.

3. Results

The graph shows the high correlation of the data with and , meaning that the resulting regression can easily account for the distribution of the results measured with a dust measuring instrument up to 2.5 μm, 87 percent. Concentration prediction obtained from a simple dust measuring device up to 2.5 microns in size using a simple regression equation was as follows: . It was found that dust concentrations of no more than 2.5 microns were found, obtained from a simple dust measuring device that measures no more than 2.5 μm with an increase of 1 μg/m3. As a result, the dust concentration of up to 2.5 μm obtained by the DustTrak 8533 was increased by 2.33 μg/m3, where is the predicted value of the particulate concentration equation of 2.5 μm obtained from simple dust measuring instruments up to 2.5 μm which are given as values from DustTrak 8533, and is the concentrations of dusts up to 2.5 μm obtained from dust detectors. Size is not more than 2.5 μm easily. A PHPM instrument for monitoring PM2.5 concentration is shown in Figure 3.

3.1. Observed and Estimated PM2.5 Concentrations

Overall, a positive linear relationship was observed between the PHPM estimates and the PM2.5 data from the DustTrak 8533 (Figure 4). The highest PM2.5 concentrations were observed in the chamber (Figure 4(a)). Ambient air PM2.5 concentrations at the three outdoor sites were similar. However, the PM2.5 was lowest at the University campus site (Figure 4(b)) and highest at rural area near a sugar cane factory (rural industrial) (Figure 4(c)). Overall, the estimates from the PHPM were lower than the data from the DustTrak 8533 (Table 3).

3.2. Correlation of PHPM Data to DustTrak 8533 Data

Considering only the outdoor ambient air sites, the correlation between PHPM and DustTrak 8533 was high, with a Pearson’s correlation coefficient values () (Table 3). The chamber data, however, show nonlinear relationship and lower correlation. In general, the values reported by the PHPM system were lower than the DustTrak 8533 estimates, indicating the PHPM device was underestimating PM2.5. Nevertheless, the PHPM system performed well with high correlation with PM2.5 concentrations in its targeted environment of ambient air. Linear regression determined the relationship of PM2.5 from the PHPM device to be

The data suggests that this linear relationship can be used to correct the PHPM data for 30-minute PM2.5 concentrations using a correction factor of 2.33 and is applicable for use in ambient air with PM2.5 concentrations in the range of 50 to 600 μg/m3. The results of the study of the relationship of the data in the study method are shown in Figures 59.

4. Discussion

This research found that the correlation between the low-cost light-scattering sensor and the higher accuracy DustTrak device was relatively high () across all observed PM2.5 concentrations in the three outdoor ambient air test sites. However, the lowest observed PM2.5 concentration was 48 μg/m3 according to the DustTrak device. Testing of multiple commercial low-cost PM2.5 sensors has demonstrated that the correlation between such low-cost devices and the DustTrak device can have poor correlation at concentrations less than 50 μg/m3 [23]. Therefore, it is possible that the PHPM device may be inaccurate at concentrations less than 50 μg/m3. However, the targeted application of the device is to increase awareness of PM2.5 concentrations during PM2.5 pollution events (generally higher than 50 μg/m3). Therefore, inaccuracy at this range does not substantially impact the intended use.

A limitation of the device is that substantial variations in conditions may affect the correction factor determined from this research. While air quality data from low-cost sensors is extremely valuable for both research and public health usage, it should be carefully evaluated, especially under unusual atmospheric or genesis conditions. In addition, the sensors will automatically alarm when the dust concentration is detected over the standard. In the dust season, people were alarmed, and they felt panic; thus, we decided to turn off the alarm. The sensors need to be used with Internet signal and also need a cloud storage for the application in order to link between dust quantity and health; anyway in some area, there is no Internet available, so the receivers needs to be installed and also a cloud storage. First, the performance of low-cost PM2.5 sensors, such as optical particle counters, can be impacted by weather conditions. High relative humidity has been shown to cause optical particle counters to overestimate PM2.5 concentrations [24]. Under high humidity conditions, hygroscopic growth of particles and reduced molecular mass of water result in an overestimation of particle mass [25]. Secondly, the varying size and composition of particles can also substantially affect the performance of low-cost PM2.5 sensors [26, 27]. The effect of differing particle composition was observed in this study, too. While the sensors performed well () in the ambient outdoor environments, the correlation was low in the chamber test using incense (). For this study, this PHPM device is more suitable for ambient air than the chamber. Thirdly, variations in the composition of atmospheric dust or meteorological conditions may also require adjustments in the computation methods for PM2.5 estimation or the use of short time periods (e.g., one hour) [24]. Thus, caution should be applied in using the PHPM sensor outside of the tested conditions. If changes in the composition or size of the PM2.5 particles, relative humidity, or desired temporal resolution occur, additional validation will be needed.

Another potential solution to the need for revised correction factors under varying conditions is to incorporate these confounding factors into the regression. While linear regression was used in this study, the accuracy of the low-cost device could be improved with multivariate regression that accounts for effects of confounding factors such as relative humidity or site-specific composition of the particulate matter [6] or nonlinear regression to account for nonlinear associations [28]. The widespread adoption of a low-cost PM2.5 sensor will help increase public awareness of current PM2.5 concentrations. The goal of increased awareness is to encourage coping activities, such as wearing a mask, [20] and pollution mitigation actions, such as using public transport [29]. Low-cost sensors can also facilitate an environment of citizen science, in which nonscientists actively participate in generating and sharing air quality data [30]. Such low-cost sensors have been used in open-source platforms to create a network of thousands of PM2.5 monitoring devices [31], in networks of community air monitoring stations [32], and to measure other pollutants besides just PM2.5 [33]. In addition, depending on the sensing capabilities of the developed and maintenance sensor probe, working on the method is to be specific and accurate for detection [7, 8].

5. Conclusion

A novel low-cost PM2.5 sensor using light-scattering technology, PHPM, was developed and validated under both the chamber and field ambient air conditions. The PHPM has a large LED display screen, to be easy to see when installing in public areas where people can be aware of the amount of dust in the area where they live. In addition, the tool can be used outdoors due to its heat-resistant metal body. The instrument is working; it will suck the air through the air intake in the sensor. There will be a laser light source acting as part of the beam to strike particles of various sizes that enters the machine by a dust measuring device that does not exceed 2.5 microns in size.

Validation against a DRX Aerosol Monitor 8533 found that the average concentration of PM2.5 for the PHPM device was using multiply with 2.33, according to linear regression. Under ambient air conditions with , the PHPM device correlated well with the DustTrak device (). However, the correlation was weaker under indoor ambient air conditions with incense as a particle source. The hope is that the PHPM device can provide an additional tool for participatory citizen science and increased public awareness about PM2.5 in Thailand. The PHPM device is inexpensive, portable, and can be charged with a power bank. Additionally, a simple measuring device for dust of less than 2.5 μm with a unique characteristic body made of heat-resistant metal can be used outdoors. The front has a large digital display of the amount of dust particles up to 2.5 μm. In the front part, there is a dust level display, no more than 2.5 μm, red, yellow, and green lights on the top, and there is a ventilation hole. The internal sensor used to read the amount of dust, the size of not more than 2.5 μm, is PMS 7003. There are components, which are composed of a heat-resistant metal body, an invention that has never been done before. (1) The front has a digital display of the amount of dust particles up to 2.5 μm. (2) In the front part, there is the level display of particulate matter up to 2.5 μm in red, yellow, and green lights is a new invention (3) with dust levels in the range of ≥50 μg/m3 red indicator light. When the dust concentration is higher than the standard value, it affects health, and there will be an alert sound immediately. The level of dust is in the range of 0-25 μg/m3 green indicator light. When the dust is concentrated, there is no health effect, and the level of dust is in the range of 26-49 μg/m3 yellow indicator light, when the dust concentration is starting to affect health. The sensor is a processor that controls the status light. A heat-resistant metal measuring instrument is built into the unit with a PMS 7003 sensor. The sensor has an air intake port attached to the PMS 7003 sensor unit. The top of the measuring instrument has several air intake holes to allow the sensor to absorb a lot of air. This is a new invention. (4) It allows the sensor to read the amount of dust particles up to 2.5 μm. (5) The sensor’s characteristics are a one-time response time of less than 1 second, integrated response time of less than or equal to 10 s charging 5 V, less than or equal to 100 mA working environment -10~+60°C, humidity 0~99% measuring range 0-500 μg/m3, and resolution 1 μg/m3 meter accuracy , @0~100 μg/m3. The limitation is that it cannot be used in environments with a dust concentration of 2.5 μm or more 500 μg/m3 and cannot be used in environments with excessive humidity and temperature.

Data Availability

The data used during the current study are available from the authors upon reasonable request.

Conflicts of Interest

The authors declare no conflict of interest.

Authors’ Contributions

P.S., P.K., and S.S., are responsible for the conceptualization; P.S., P.K., W.L., and R.S. for methodology; W.L. for software; P.S., P.K., W.L., R.S., N.M., and J.R. for validation; P.K., W.L., and R.S. for formal analysis; W.L., R.S., and C.J. for resources; P.S. and S.S. for writing—original draft preparation; P.S. and S.S. for writing—review and editing; P.K. for visualization; and P.S. for funding acquisition. All authors have read and agreed to the published version of the manuscript.

Acknowledgments

We express our sincere appreciation to the Occupational Health Safety and Environmental Epidemiology Group (OHSEE) at the Faculty of Public Health, Khon Kaen University for data collection and operator. This research was funded by research program funding from Khon Kaen University (grant number: RP64006).