Fault Detection Method of Medical Equipment Based on Multi-Index Electrical Performance Parameters
Table 1
Current mean value of a certain type of monitor under various faults.
| | Status name | Value name | Normal | ECG fault | Mainboard fault | Display screen fault | Main cable fault | Air pump fault |
| Average shutdown current | 0.140 A | 0.137 A | 0.051 A | 0.122 A | 0.122 A | 0.141 A | Average stand by current | 0.205 A | 0.188 A | 0.139 A | 0.188 A | / | 0.196 A | Threshold range | (−6, 6 mA) | (−5, 5 mA) | (−5, 5 mA) | (−5, 5 mA) | (−5, 5 mA) | (−5, 5 mA) |
|
|