Research Article

Fault Detection Method of Medical Equipment Based on Multi-Index Electrical Performance Parameters

Table 8

Judgment accuracy rate of various indicators under different operating statuses.

StatusesNormal (%)ECG circuit module fault (%)Motherboard fault (%)Screen
malfunction (%)
Main cable
malfunction (%)
Air pump
fault (%)

Current only10048.110085.374.230.5
Voltage only10095.389.238.226.55
Power only10083.780.162.981.544.7
Multi-indicators10099.5610096.4797.393.55