Research Article
Fault Detection Method of Medical Equipment Based on Multi-Index Electrical Performance Parameters
Table 8
Judgment accuracy rate of various indicators under different operating statuses.
| Statuses | Normal (%) | ECG circuit module fault (%) | Motherboard fault (%) | Screen malfunction (%) | Main cable malfunction (%) | Air pump fault (%) |
| Current only | 100 | 48.1 | 100 | 85.3 | 74.2 | 30.5 | Voltage only | 100 | 95.3 | 89.2 | 38.2 | 26.5 | 5 | Power only | 100 | 83.7 | 80.1 | 62.9 | 81.5 | 44.7 | Multi-indicators | 100 | 99.56 | 100 | 96.47 | 97.3 | 93.55 |
|
|