Research Article

Comparison of Performance of Photodiodes with Different Active Areas Using Acrylic and Quartz Cuvettes for Spectrophotometry in Direct Measurements of Glucose in Water and Human Blood Plasma by Optical Means Using Near-Infrared

Table 3

Downward voltage pattern using acrylic and quartz cuvettes.

Glucose concentration (mg/dl)Voltage using acrylic cuvette (V)Decrease using acrylic cuvette (mV)Voltage using quartz cuvette (V)Decrease using quartz cuvette (mV)

02.233NA3.253NA
1002.22493.2458
2002.212123.23510
3002.20573.22411
4002.19873.21212
5002.18993.2057
6002.18633.19114
7002.18243.1829
8002.169133.16913
9002.16633.15514
10002.15883.14411