Research Article

Comparison of Performance of Photodiodes with Different Active Areas Using Acrylic and Quartz Cuvettes for Spectrophotometry in Direct Measurements of Glucose in Water and Human Blood Plasma by Optical Means Using Near-Infrared

Table 5

Glucose measurement using active area of 1.5, 1, 0.5, and 0.06 mm.

Glucose concentration (mg/dl)Voltage using 1.5 mm active area (V)Decrease using 1.5 mm active area (mV)Voltage using 1 mm active area (V)Decrease using 1 mm active area (mV)Voltage using 0.5 mm active area (mV)Decrease using 0.5 mm active area (mV)Voltage using 0.06 mm active area (mV)Decrease using 0.06 mm active area (mV)

03.253NA1.391NA443NA20NA
1003.24581.3865441219.90.1
2003.235101.3833437419.70.2
3003.224111.3785436119.70
4003.212121.3717434219.70
5003.20571.3674431319.60.1
6003.191141.3625430119.60
7003.18291.362428219.50.1
8003.169131.3564427119.50
9003.155141.3533426119.40.1
10003.144111.353424219.40