Research Article
Mechanism and Modeling of Contaminant Accumulation on Hot-Film Air Flow Sensor
Table 8
Flow rate measured by No.6 and No.7 sensors.
| Test time (h) | Reference value | No.6 sensor | No.7 sensor | (L/min) | Flow rate (L/min) | Decrease amplitude of accuracy (%) | Flow rate (L/min) | Decrease amplitude of accuracy (%) |
| 300 | 15 | 14.4 | 4% | 14.0 | 6.667% | 600 | 15 | 13.9 | 7.333% | 13.3 | 11.333% | 900 | 15 | 13.6 | 9.333% | 12.6 | 16% | 1200 | 15 | 13.2 | 12% | 11.9 | 20.667% |
|
|