Research Article

Exploring the Clinical Benefits of Mixed-Reality Technology for Breast Lumpectomy

Table 2

Fusion error measured on selected 5 marker points on the 10 tested patients using 3D printing breast models (unit: mm).

 Error of marker point 1Error of marker point 2Error of marker point 3Error of marker point 4Error of marker point 5

Data11.451.831.960.260.61
Data20.900.820.421.193.34
Data30.130.870.390.332.43
Data40.800.401.471.611.06
Data50.371.171.670.950.96
Data60.290.831.531.050.95
Data71.131.660.820.493.08
Data81.831.570.861.051.94
Data91.080.550.271.000.33
Data101.810.921.351.381.05
Average0.981.061.070.931.57
SD0.600.480.590.441.05

SD, standard deviation.