Implementation of Bioelectrical Impedance Measuring Instrument Based on Embedded System

Table 6

Difference error between the bioelectrical impedance of human left and right arm at multiple frequencies using the developed and standard devices.

Frequency (kHz)Impedance (Ω)
Mean value
Left armRight arm
Standard deviceBIA instrumentDiff./error (%)Standard deviceBIA instrumentDiff./error (%)

5433.2399.37.83437.2390.410.70
50378.6357.65.55358.3351.71.84
100356.3339.64.69339.2334.31.44
150348.8330.85.16330.2326.21.21
200336.4321.44.46323.4317.51.82
250326.5318.72.39320.7314.32.00
300319.2315.21.25311.4311.60.06
400316.1312.11.27304.4307.40.99