Implementation of Bioelectrical Impedance Measuring Instrument Based on Embedded System
Table 6
Difference error between the bioelectrical impedance of human left and right arm at multiple frequencies using the developed and standard devices.
| Frequency (kHz) | Impedance (Ω) Mean value | Left arm | Right arm | Standard device | BIA instrument | Diff./error (%) | Standard device | BIA instrument | Diff./error (%) |
| 5 | 433.2 | 399.3 | 7.83 | 437.2 | 390.4 | 10.70 | 50 | 378.6 | 357.6 | 5.55 | 358.3 | 351.7 | 1.84 | 100 | 356.3 | 339.6 | 4.69 | 339.2 | 334.3 | 1.44 | 150 | 348.8 | 330.8 | 5.16 | 330.2 | 326.2 | 1.21 | 200 | 336.4 | 321.4 | 4.46 | 323.4 | 317.5 | 1.82 | 250 | 326.5 | 318.7 | 2.39 | 320.7 | 314.3 | 2.00 | 300 | 319.2 | 315.2 | 1.25 | 311.4 | 311.6 | 0.06 | 400 | 316.1 | 312.1 | 1.27 | 304.4 | 307.4 | 0.99 |
|
|