Research Article

Validation of Two Activity Monitors in Slow and Fast Walking Hospitalized Patients

Table 9

Relative and absolute reliability of steps, walking, and sedentary time assessed with ActivPAL3™, SENS motion®, and direct observations in patients following acute high-risk abdominal surgery ().

ComparisonMean (SD) of both assessments (sec)Mean (SD) difference (sec) value for differenceICC2.1 (95% CI)SEM (SEM%)MDC95 (MDC%)

Steps
ActivPAL vs. observed197.5 (80.2)-88.1 (52.3)<0.0010.529 (-0.105-0.868)55.0 (28)152.5 (77)
SENS vs. observed206.9 (75.3)-69.4 (32.3)<0.0010.653 (-0.075-0.920)44.4 (21)123.0 (59)
ActivPAL vs. SENS162.8 (87.5)-18.7 (33.2)0.090.916 (0.699-0.977)25.4 (16)70.4 (43)
Walking time
ActivPAL vs. observed138.7 (38.1)-64.2 (51.7)0.0020.193 (-0.135-0.612)34.2 (25)94.8 (68)
SENS vs. observed158.9 (20.4)-23.9 (18.0)0.0010.435 (-0.121-0.812)15.3 (10)42.4 (27)
ActivPAL vs. SENS126.7 (34.9)-40.4 (57.6)0.0420.143 (-0.264-0.610)32.3 (25)89.5 (71)
Sedentary time
ActivPAL vs. observed195.9 (21.8)-9.8 (13.8)0.0400.760 (0.263-0.932)10.7 (5)29.7 (15)
SENS vs. observed196.5 (21.6)-8.6 (9.4)0.0120.853 (0.309-0.964)8.3 (4)22.2 (11)
ActivPAL vs. SENS191.6 (23.9)-1.2 (14.7)0.80.840 (0.506-0.954)9.6 (5)26.6 (14)

ICC: intraclass correlation coefficient; SEM: standard error of measurement (); MDC: minimal detectable change (, ). SEM and MDC indicate measurement error (in sec) at group and individual levels, respectively. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.