Research Article
MR Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types
| No. | Name | Score signal intensity | Score lumen narrowing | Score lumen homogeneity | Overall Score | Overall Evaluation |
| 1 | SAXX Small | 1 | 1 | 1 | 3 | Poor | 2 | SelfX Xpert | 3 | 2 | 2 | 7 | Intermediate | 3 | Palmaz Genesis | 1 | 1 | 1 | 3 | Poor | 4 | Absolute | 3 | 2 | 3 | 8 | Good | 5 | AccuLink Carotid | 2 | 2 | 2 | 6 | Intermediate | 6 | Express Vascular LD | 1 | 1 | 1 | 3 | Poor | 7 | OmniLink 0.018 | 2 | 2 | 2 | 6 | Intermediate | 8 | OmniLink 0.035 | 2 | 2 | 2 | 6 | Intermediate | 9 | Palmaz Corinthian IQ | 1 | 1 | 1 | 3 | Poor | 10 | Renal109 | 3 | 3 | 2 | 7 | Good | 11 | Renal137 | 3 | 2 | 3 | 8 | Good | 12 | Sentinol | 2 | 3 | 3 | 8 | Good | 13 | Symphony | 1 | 2 | 2 | 5 | Intermediate | 14 | Vascuflex SE | 2 | 3 | 3 | 8 | Good | 15 | Wallstent Uni | 3 | 2 | 3 | 8 | Good | 16 | Zilver | 3 | 2 | 3 | 8 | Good | 17 | Evo | 2 | 3 | 2 | 7 | Intermediate | 18 | RxCarotid | 1 | 2 | 1 | 4 | Poor | 19 | SAXX Large | 1 | 1 | 1 | 3 | Poor | 20 | Evo Target | 2 | 3 | 2 | 7 | Intermediate | 21 | Wallgraft Endoprothesis | 2 | 3 | 2 | 7 | Intermediate | 22 | CP Stent | 1 | 3 | 2 | 6 | Intermediate |
|
|