A Review of Doses for Dental Imaging in 2010–2020 and Development of a Web Dose Calculator

Table 1

Technical parameters for intraoral, panoramic, and cone-beam computed tomography collected from literature, which were used for Monte Carlo simulations conducted using MC-GPU to estimate dose distribution within the head.

ParametersIntraoralPanoramicCone-beam computed tomography

X-ray energy (kVp)607390
Filtration (mm Al)3.52.52.8
FOV (cm2)4 × 310 × 0.210 × 10
SRD1 (cm)753550
Rotation angle per view (degree)02240360
Number of views per acquisition19153360
Number of X-rays per simulated view10115 × 1075 × 109
Total simulation time (min)333143760

1Source-to-rotation axis distance. 2The rectangular field was rotated 30° cranially. 3Simulation run in an NVIDIA GeForce GTX 1080 GPU.