Research Article

Effect of Amalgam Restorations and Operation Parameters on Diagnostic Accuracy of Caries Detection Using Cone Beam Computed Tomography: An In Vitro Study

Table 2

The mean sensitivity, specificity, and AUC scores for all three readers combined for approximal caries detection.

ModeSensitivity (true positive rate)Specificity (1—false positive rate)AUC (area under ROC curve)95% CI for AUC
Without ARaWith ARbWithout ARaWith ARbWithout ARaWith ARbWithout ARaWith ARb

Imaging Mode 1
90kVp, 8mA, 75μm
0.9720.9660.9730.5380.9760.8960.957-0.9940.849-0.942
Imaging Mode 2
80kVp, 7mA, 75μm
0.8750.9160.3750.2560.7710.6500.704-0.8390.559-0.742
Imaging Mode 3
80kVp, 7mA, 200μm
0.9300.8330.8120.4870.9450.8230.911-0.9790.755-0.891
Imaging Mode 4
90kVp, 8mA, 200μm
0.9710.9830.8750.6920.9660.9340.944-0.9860.906-0.959

aBlocks for teeth without amalgam restorations; bBlocks for teeth with amalgam restorations.