Review Article

Recent Advances of Hepatitis B Detection towards Paper-Based Analytical Devices

Table 6

Paper-based analytical devices (PADs) with various detections for HBV detection.

BiomarkerDetectionLinearity rangeDetection limitRef.

HBsAgECL34.2 pg·mL−1 to 34.2 ng·mL−134.2 pg·mL−1[46]
HBsAgColorimetric1–50 ng·mL−10.37 ng·mL−1[44]
HBV DNAColorimetricNS100 IU·mL−1[45]
HBV DNAASV0–500 pM and 0–1.5 nM85 pM[47]
HBV DNADPV50–100 pM1.45 pM[48]

NS, not specified due to qualitative analysis.