Mean value and homogeneity of magnetic flux density with a current in the sweep coil evaluated for 2 cm DSV. The X-axis is in the direction where the main magnetic field penetrates the two pole faces. The Y-axis is a horizontal plane between two pole faces.

Sweep currentMethodX-axisY-axis
Mean (mT)Homogeneity (ppm)Mean (mT)Homogeneity (ppm)

−4 AMeasurement42.9869843.053,484
FEM42.041,28842.06583
+4 AMeasurement45.862,18145.862,617
FEM44.8840044.89268