Research Article

Odd-Leg Birdcages for Geometric Decoupling in Multinuclear Imaging and Spectroscopy

Table 2

Double-tuned birdcage bench measurements.

Measurement1HSingle-Tuned23NaSingle-Tuned1HDouble-Tuned23NaDouble-Tuned1HDouble-Tuned Active Detuning23NaDouble-Tuned Active Detuning

Q (AU)344337301326233284
S11 (dB)−29.7−34.4−41.6−30.7−37.5−32.0
S21 (dB)N/AN/A−21.2−18.3−16.6−22.3