DYNEELAX Robotic Arthrometer Reliability and Feasibility on Healthy and Anterior Cruciate Ligament Injured/Reconstructed Persons

Table 2

Mean, standard deviation, ICC, SEM, and MDC results for the test-retest.

Mean ± SDICC (95% CI)SEMMDCMDC (%)

(mm)4.9 ± 1.50.96 (0.94 to 0.97)0.300.8316.8
PCa (μm/N)36.4 ± 15.30.95 (0.93 to 0.97)3.329.2125.3
SCa (μm/N)30.3 ± 15.30.91 (0.87 to 0.94)2.005.5318.3
(°/5 Nm)10.9 ± 4.40.94 (0.91 to 0.96)1.082.9827.3
(°/5 Nm)11.3 ± 4.90.96 (0.93 to 0.97)1.042.8825.5
2.8 ± 0.90.91 (0.87 to 0.94)0.260.7225.6
2.8 ± 1.00.93 (0.89 to 0.95)0.280.7627.7

SD, standard deviation; ICC (95% CI), intraclass correlation coefficient with 95% confidence interval; SEM, standard error of measurement; MDC, minimum detectable change; , translation at 150 N; PCa, primary compliance; SCa, secondary compliance; , internal rotation at 5 Nm; , external rotation at 5 Nm; , internal rotation slope; , external rotation slope.