Comparison of Rhenium and Iodine as Contrast Agents in X-Ray Imaging
Table 3
Settings for digital radiography.
| | Exposure ID | X-ray tube potential (kVp) | Amount of additional filtration (mm of Cu) | Current (mA) |
| | X12 | 50 | 0 | 1.25 | | X13 | 50 | 0.1 | 2.5 | | X14 | 50 | 0.3 | 10 | | X15 | 50 | 0.5 | 36 | | X16 | 50 | 1 | 71 | | X17 | 50 | 1.5 | 71 | | X18 | 81 | 0 | 1.25 | | X19 | 81 | 0.1 | 1.6 | | X20 | 81 | 0.3 | 1.6 | | X21 | 81 | 0.5 | 2.2 | | X22 | 81 | 1 | 4 | | X23 | 81 | 1.5 | 4 | | X24 | 121 | 0 | 1.25 | | X25 | 121 | 0.1 | 1.25 | | X26 | 121 | 0.3 | 1.25 | | X27 | 121 | 0.5 | 1.25 | | X28 | 121 | 1 | 1.25 | | X29 | 121 | 1.5 | 1.25 |
|
|
The Re- and I-XCA were imaged at 50, 81, and 121 kVp using a clinical X-ray camera. The current was chosen to avoid saturation of the detector.
|