Comparison of Rhenium and Iodine as Contrast Agents in X-Ray Imaging
Table 4
Measurement of the incident air kerma and the mean energy of the X-ray beam.
| Scan ID | X-ray tube potential (kVp) | Amount of additional filtration (mm of Cu) | Incident air kerma (Gy) | Mean energy (keV) |
| X17 | 50 | 0 | 14.9 + 0.1 | 28.5 | X18 | 50 | 0.1 | 10.5 | 36.2 | X19 | 50 | 0.3 | 11.2 | 39.6 | X20 | 50 | 0.5 | 15.4 | 41.4 | X21 | 50 | 1 | 4.2 | 43.9 | X22 | 50 | 1.5 | 1.3 | 45.1 | X23 | 81 | 0 | 49.3 + 0.2 | 43.1 | X24 | 81 | 0.1 | 31.9 + 0.6 | 48.1 | X25 | 81 | 0.3 | 15.4 + 0.1 | 53.4 | X26 | 81 | 0.5 | 12.5 | 56.5 | X27 | 81 | 1 | 9.3 | 61.0 | X28 | 81 | 1.5 | 21.2 + 0.1 | 63.7 | X29 | 121 | 0 | 114.9 + 0.2 | 53.9 | X30 | 121 | 0.1 | 74.0 + 0.2 | 59.7 | X31 | 121 | 0.3 | 46.0 + 0.1 | 65.9 | X32 | 121 | 0.5 | 33.4 + 0.1 | 69.8 | X33 | 121 | 1 | 19.7 + 0.1 | 76.1 | X34 | 121 | 1.5 | 11.7 + 0.1 | 80.5 |
|
|
The incident air kerma is reported as the mean of three independent measurements. The standard deviation is zero in cases where it is not reported.
|