Table 1

OBI safety and functional test items.

The test itemsA detailed descriptionResults

The door chainIf the shielding door of the accelerator room is not closed, we use the ray tube lockCannot send X-rays
Warning lightsAll warning lights should be kept on during X-ray transmissionThe warning light is on
Warning soundWhen X-ray is transmitted, OBI manual control box should sound a warning soundWarning tone appears
Collision detection and interlockingIn the process of KVS and KVD motion intentional collision, check whether the collision chain appearsMotion stops, warning lights, and warning sounds appear
OBI manual control box buttonRelease the manual box button during KVS and KVD motionThe movement stops
Injection line preheatingStep on the foot pedal to warm up the tube in perspective mode. The default preset parameters are 75 kV voltage, 50 mA current, and 32 ms exposure timeExecutable operation
FunctionalExamination data are transmitted from the treatment planning system to the treatment terminal, OBI workstation, and mechanical systemExecutable operation