OBI safety and functional test items.
The test items
A detailed description
Results
The door chain
If the shielding door of the accelerator room is not closed, we use the ray tube lock
Cannot send X-rays
Warning lights
All warning lights should be kept on during X-ray transmission
The warning light is on
Warning sound
When X-ray is transmitted, OBI manual control box should sound a warning sound
Warning tone appears
Collision detection and interlocking
In the process of KVS and KVD motion intentional collision, check whether the collision chain appears
Motion stops, warning lights, and warning sounds appear
OBI manual control box button
Release the manual box button during KVS and KVD motion
The movement stops
Injection line preheating
Step on the foot pedal to warm up the tube in perspective mode. The default preset parameters are 75 kV voltage, 50 mA current, and 32 ms exposure time
Executable operation
Functional
Examination data are transmitted from the treatment planning system to the treatment terminal, OBI workstation, and mechanical system