Comparison of the Accuracy of Fixture-Level Implant Impression Making with Different Splinting Techniques

Table 1

Comparison of impression accuracy.

Splinted, impression after the setting time (4 min)NMeanStandard deviationOne-sample t-test (value = 19.164)
Mean difference95% confidence interval for the difference value
Lower boundUpper bound

Distance1219.14170.02907−0.0223−0.0408−0.00390.022

17 minutes, cutting and splinting againNMeanStandard deviationOne-sample t-test (value = 19.164)
Mean difference95% confidence interval for the difference value
Lower boundUpper bound

Distance1219.15410.03949−0.00992−0.0350.01520.403

24 hours, cutting and splinting againNMeanStandard deviationOne-sample t-test (value = 19.164)
Mean difference95% confidence interval for the difference value
Lower boundUpper bound

Distance1219.15900.03369−0.00500−0.02640.01640.617