Effect of the Field of View Size on CBCT Artifacts Caused by the Presence of Metal Objects in the Exomass

Table 1

Descriptive findings regarding the mean GV based on the size of FOV and the type and number of metal objects (n = 10).

GroupNumberMinimumMaximumMeanStd. errorStd. deviationVariance

6  6-Co-Cr-210.00748.87775.59767.962.838.9580.10
8  8-Co-Cr-210.00610.40828.70716.1422.4871.105055.31
6  6-amalgam-210.00451.50609.20545.5017.8856.533195.64
8  8-amalgam-210.00589.40657.50629.407.4023.40547.54
6  6-amalgam-110.00470.60653.80569.8221.1166.754455.92
8  8-amalgam-110.00358.30689.20502.0431.63100.0210004.02
6  6-titanium-210.00841.70848.27845.650.591.873.50
8  8-titanium-210.00954.90971.60962.781.805.6832.31
6  6-titanium-110.00705.80790.50763.389.1028.78828.35
8  8-titanium-110.00933.50940.30937.420.702.214.88
6  6-control-110.00884.80892.30889.200.822.606.75
6  6-control-210.00888.70892.10890.760.351.111.23
8  8-control-210.00951.80960.90955.380.772.445.94
8  8-control-110.00968.30974.00971.990.571.803.24
6  6-Co-Cr-110.00860.10864.30862.100.411.311.71
8  8-Co-Cr-110.00915.50919.00917.520.300.950.91