Research Article
Heavy Metal Residue (As, Cd, Hg, and Pb) in Hen Eggs after Applying Different Cooking Methods
Table 1
Conditions of ICP-MS apparatus for determining As, Pb, Cd, and Hg in hen eggs.
| Parameter | Value |
| Radiofrequency | 1200 W (40 Mhz) | Plasma gas (argon) flow | 16 l/min | Nebulizer gas (argon) flow | 1 l/min | Read delay and analysis speeding | 30 s | Wash | 60 s | Wash speeding | 30 rpm | Dwell time | 50 ms | Resulting/amu 10% peak | 0.7 | Integration time | 3.5 | Linear working (total element) (ppb) | 0.053 | Precision %RSD | 1.3 | Addition/recovery | 93-103 | Repetition | 3 | Limit of detection (LOD) (μg kg-1) | 0.00033 | Limit of quantification (LOQ) (μg kg-1) | 0.001 |
|
|